qPCR Efficiency Calculator

Slope of Ct plotted against log₁₀ of the starting quantity. It is negative; about −3.32 means perfect doubling.
Biology Updated 29 Oct 2026

Finds the amplification efficiency of a real-time (quantitative) PCR assay. Enter the slope of your standard curve, or paste the Ct values of a serial dilution and let the calculator fit the line and report R². It works in reverse too: turn an efficiency or amplification factor back into the expected slope. Results show the efficiency in percent, the amplification factor per cycle and whether the assay falls in the usual 90–110% window.

  • Ct values are in the log-linear range of the assay and the dilution series is exact.
  • Efficiency is constant across cycles and across the dilution range.
  • In standard-curve mode every well is weighted equally; outlier removal is left to you.
  • Slopes outside −20 to −1 are rejected because they correspond to physically impossible or meaningless efficiencies.

Formula

Efficiency E = 10^(−1/slope) − 1 (× 100 for %). Amplification factor = 10^(−1/slope) = 1 + E. Slope = −1 ÷ log₁₀(1 + E).
  • The slope comes from a standard curve: Ct of each dilution plotted on the y-axis against log₁₀ of its starting quantity on the x-axis. More template means fewer cycles, so the slope is negative.
  • If the product doubles every cycle (amplification factor 2, efficiency 100%), a 10-fold dilution delays Ct by log₂(10) = 3.322 cycles, so the slope is −3.322.
  • In standard-curve mode the line is fitted by ordinary least squares through every replicate well, with x = log₁₀(top quantity) − point index × log₁₀(dilution factor). R² = Sxy² ÷ (Sxx × Syy).
  • The usual acceptance window is 90–110% efficiency, which corresponds to slopes of about −3.59 to −3.10.
  • Results are rounded for display: efficiency to 2 decimals, amplification factor to 4 decimals and slope to 3 decimals.

Standard-curve slope of −3.35

Inputs
  • Calculate from: slope
  • Slope: -3.35
  • Quantity in the most concentrated standard:

Amplification factor = 10^(−1/−3.35) = 10^0.2985 = 1.9884. Efficiency = 1.9884 − 1 = 0.9884, or 98.84%. A 10-fold dilution shifts Ct by 3.35 cycles, and the assay is inside the 90–110% window.

Sources & references

Frequently asked questions

What is a good qPCR efficiency?
Most laboratories and assay suppliers treat 90–110% as acceptable, with 95–105% preferred for comparative (ΔΔCt) quantification; the MIQE guidelines ask you to report the efficiency, slope and R² of the standard curve. 90–110% corresponds to a slope between roughly −3.6 and −3.1, and the fit should have R² of at least 0.98.
Why can efficiency be above 100%?
True amplification cannot exceed doubling per cycle, so values above 100% point to an artefact. Common causes are polymerase inhibitors in the concentrated standards (which delay their Ct and flatten the curve), pipetting errors in the dilution series, primer dimers or non-specific products detected by SYBR Green, and too few dilution points.
How do I make a standard curve?
Make a serial dilution of template, usually 5 to 7 points of 10-fold (or 2- to 5-fold for a narrow range), run each point in triplicate, and plot the Ct values against log₁₀ of the input amount. Drop points that are in the noisy late-Ct region or show inhibition, then read the slope of the straight line.
What does the amplification factor mean?
It is how many times the product multiplies in one cycle: 2.0 for perfect doubling, 1.9 for 90% efficiency. Efficiency-corrected methods such as the Pfaffl method use this factor instead of assuming 2 when comparing genes.
Does the quantity unit or top concentration matter?
No for efficiency. Changing the units or the top quantity shifts the whole line left or right without changing its slope. It only changes the y-intercept, which you need when you want to read absolute copy numbers off the curve.

Related calculators

DNA Copy Number Calculator Converts a DNA or RNA concentration into the number of template molecules per microlitre, or a mass of DNA into a total number of copies. Enter the concentration (or mass), the template length and whether the template is double-stranded DNA, single-stranded DNA or RNA; the calculator applies Avogadro's constant and the average mass per base pair. An optional dilution helper tells you how much stock and water to mix for a target copies/µL, and the PCR section estimates copies after a number of cycles, or the cycles needed to reach a target, at any amplification efficiency. DNA Concentration Calculator Turns a UV spectrophotometer reading into a nucleic acid concentration. Enter the absorbance at 260 nm, the cuvette path length and any dilution factor, and pick double-stranded DNA (50 µg/mL per A₂₆₀), single-stranded DNA (33), RNA (40), your own conversion factor, or a custom oligonucleotide. For an oligo, paste its sequence and the calculator works out the molecular weight and the nearest-neighbour extinction coefficient, or type both values in. Optional A₂₈₀ and A₂₃₀ readings give the purity ratios, and a sample volume gives the total yield. Annealing Temperature Calculator Finds the optimal PCR annealing temperature with the Rychlik formula Ta = 0.3 × Tm(primer) + 0.7 × Tm(product) − 14.9. Enter the melting temperature of the less stable primer and of the PCR product in °C, °F or K, or paste both primer sequences and let the calculator work out each primer's Tm (nearest-neighbour, salt-adjusted or basic method) and the product Tm from its length and GC content at your salt concentration. It also flags primer pairs whose melting temperatures differ by more than 5 °C. Cell Dilution Calculator Plan a cell suspension dilution with C1 × V1 = C2 × V2. Enter any three of the stock concentration, the volume of stock to take, the target concentration and the final volume, each with its own unit (µL, mL, cL or L), and the calculator finds the fourth. It also reports the dilution factor, how much medium or buffer to add, the total number of cells, and a serial dilution plan when the stock volume is too small to pipette accurately.